APiC Ray Fluoroscopy System

  • APiC Ray Fluoroscopy System
  • APiC Ray Fluoroscopy System
  • APiC Ray Fluoroscopy System
  • APiC Ray Fluoroscopy System
APiC Ray Fluoroscopy System

APiC Ray Fluoroscopy System

Description APiC EPS RF 50/80 EPS 50/80 is an Advanced High Frequency X-ray Generator designed with a unique reliable and modular approach. This un... Read More
  1. APiC EPS RF 50/80
EPS 50/80 is an Advanced High Frequency X-ray Generator designed with a unique reliable and modular approach. This unique conception delivers impressive kV waveform resulting in superior image quality and reduced soft radiation. Enabling limitless capabilities; the modular technology enables more features and power to be added when needed using plug-in modules quickly inserted to generator main core.
Features: 
- Extremely sharp kV waveform: Unparalleled rise and fall times in the industry resulting in superior image quality and reduced soft radiation.
- Exceptional High efficiency: With efficiency 92% resulting in reduced power consumption and increased equipment reliability
- Anatomical Programming Radiography (APR): Pre-programmed technique chart consisting of up to 2000 fully customizable exams, stored in the Console.
- Automatic Exposure Control (AEC): Six clinically significant optical density corrections are provided with the AEC mode, varying from the standardized optical density. Compensation ranges from - 80 % up to+ 130 % of photo timing level for 12 stations be tween 42 kV and 133 kV.
- Automatic Brightness Stabilizer (ABS) : allows optimum
Fluoro image stabilization level regardless of the
patient size. The kV is adjusted as a function of the
real time feedback of the Fluoro beam provided
by a photomultiplier tube, a photodiode or flat panel
detector.
- Fluoroscopic Entrance Exposure Rates:
The generator supports the use of a two-level
Fluoro footswitch or two separate footswitches to
be supplied by the installer. Each level has its own
Fluoro exposure entrance rate adjustment.
- Source-To-Patient Distance Tracking Mode (optional):
The generator can adjust the Fluoro beam to compensate
for the source to image distance (SPD) when the gantry position varies during Fluoro. This provides constant Fluoro
surface dose along the full SPD span.
- Single or double X-ray tubes according to customer requirements
- Integrated dual speed anode drive
- High performance radiography mode available: 
12,500 mAs & 63-s radiation time 
- Integrate PC tools 

Fluoroscopy Specifications :

. Continuous Fluoroscopy mode


kV   40 to 125 kV, in one-kV steps
mA   0.2 to 12.5 mA with 12 steps with 25% increment
High performance fluoroscopy feature enables 20 mA at 140 kV

. Pulsed fluoroscopy mode (optional)


kV 40 to 125 kV , in one-kV steps
mA 16 to 80 mA with 12 steps with 25% increment
Pulse width 5.0 to 100 ms
Frame rate 0.25 up to fps
Pulsed Fluoroscopy Operation Modes
o Normal mode dose     Fluoroscopy at 30 (25) pulses / second.
o Dose reduction           Reduced the total X-ray dose by decreasing the pulse rate.
o Image enhancement  Improved the image contrast at lower Fluoro speeds



II. APiC Remo 20/80

Diagnostic table Specifications

 Vertical movement range  Head 450 mm / End: 130 mm Left-right± 110 mm

 Tilting range  -40 ~ + 90 (-25 ~ +90 optional)
 Tabletop Load payload  200 kg
 Grid  N103, r10
 Movement of Spot Film device   600 mm
 Tabletop to spot film device  45 mm
 Radiation field adjustment  Rectangular size adjusted automatically in longitudinal and   lateral   direction up to the maximum size circumscribed by the   input plan of 1-1 / 2-2 / 3-3